SpectroLink: Python interface for GoyaLab spectrometers

SpectroLink is the Python interface of SpectroLab. With the spectroLink.py module, a Python script drives an IndiGo spectrometer through a running SpectroLab instance: exposure, gain, light sources, captures, device values and colorimetry. This page covers installation, a first script, the full API, ready-to-use examples and the packet protocol.

User script + spectroLink.py  <-- TCP 127.0.0.1:65357 -->  SpectroLab  <-- USB -->  IndiGo

SpectroLink exists in two generations. Differences are flagged throughout this page:

  • V2 – SpectroLab 2.x: one device, the one connected in the SpectroLab window.
  • V3 – SpectroLab 3.x: several devices, addressed individually.

V2 / V3 – The V2 and V3 spectroLink.py files are not interchangeable. Always use the file installed with the SpectroLab version you are running.

Contents: Requirements · Getting started · Python API · Examples · V2 / V3 summary · Protocol reference

1. Requirements and files

  • Windows 10 / 11, 64-bit.
  • SpectroLab running on the same computer, with or without its window (see 2.1). Download SpectroLab.
  • For an external interpreter: Python 3 with numpy (matplotlib for the examples).

The files are installed with SpectroLab in <SpectroLab>\utils\python (by default C:\Program Files (x86)\GoyaLab\SpectroLab). Copy spectroLink.py next to your script.

File Content
spectroLink.py Interface module
pyExample.py Usage examples: captures, sources, device values, colorimetry
test_headless.py Minimal capture test
internallink.py, internal\ Used by SpectroLab’s internal math pipeline – not for user scripts

2. Getting started

2.1 Start SpectroLab

SpectroLink needs a running SpectroLab instance, with its window or as a background process (no GUI):

cd "C:\Program Files (x86)\GoyaLab\SpectroLab"
start SpectroLab.exe --nogui
  • Only one SpectroLab instance runs at a time.
  • Wait for the end of initialisation (device detection, EEPROM read) before the first command: 3 to 20 s depending on the configuration.
  • Stop a background instance with taskkill /f /im SpectroLab.exe.

V3 – V3 also accepts -ng. Starting the GUI while a background instance is running closes the background instance.

2.2 First script

Connect, set exposure and gain, and capture one spectrum. The helper wait_spectrum() is reused in the examples below. This script runs unchanged on V2 and V3.

import time
import numpy as np
from spectroLink import *

def wait_spectrum(link, timeout_s=5.0):
    """Returns the next spectrum, or None after timeout_s."""
    t_end = time.time() + timeout_s
    while time.time() < t_end:
        if link.available() > 0:
            pkt = link.read_packet()
            if pkt.opcode() == SpectrolabOpcode.OPCODE_RECEIVED_SPECTRUM:
                return pkt.to_spectrum()
        else:
            time.sleep(0.02)
    return None

link = SpectroLink()
link.flush()
link.set_expo_ms(100)
link.set_gain_mdB(2000)                 # 2 dB
link.trigger_spectrum_capture()
s = wait_spectrum(link)
if s:
    print(s.length_points, "points,", s.X[0], "to", s.X[-1], "nm")

2.3 Run a script from SpectroLab

A script can also be run from SpectroLab’s Python module. SpectroLab runs it through pyLink.exe, an embedded Python with numpy, scipy and matplotlib. The script must define spectrolab_run(): pyLink imports the file and calls this function. print() output appears in the module console.

from spectroLink import *

def spectrolab_run():                    # entry point called by pyLink.exe
    link = SpectroLink()
    link.trigger_spectrum_capture()
    s = wait_spectrum(link)              # as defined in 2.2
    print("max:", max(s.Y))

V2 / V3 – In V2, with an external interpreter, the line at the top of spectroLink.py that makes stdout unbuffered fails in consoles without a file descriptor (pythonw, some IDEs). Remove it in that case. V3 handles this case.

3. Python API

Methods of the SpectroLink class. In V3, every command method takes an extra keyword argument targetSPNIAddress (default: the device selected in SpectroLab, see 3.5).

3.1 Connection and reception

Method Description V2 / V3
SpectroLink() Connects to SpectroLab. Raises ConnectionRefusedError if SpectroLab is not running.
flush() Clears the received packets.
available() Reads the socket and returns the number of complete packets waiting. Must be polled.
read_packet() Returns the oldest waiting packet. Empty queue: V2 [], V3 None
await_packet(timeout_s) Waits for the next packet, polling every 100 ms. V3 only
read_sernum(addr) Returns the serial number of a device, or "" after 0.5 s. V3 only
  • Commands are asynchronous and not acknowledged. Only spectra, ADC data, device values and colorimetry results come back.
  • Every connected script receives every packet, including spectra captured from the GUI: always test opcode().
  • A spectrum spans several socket reads: keep calling available() until it returns a non-zero value.

3.2 Acquisition settings

Method Description
set_expo_ms(expo_ms) Exposure time in ms (float).
set_gain_mdB(gain_mdB) Gain in millidecibels: 2 dB = 2000.
set_gain_lin(gain_lin) Linear gain, sent as 1000·log10(gain_lin) mdB.

Settings stay active until changed, including for captures started from the GUI.

3.3 Captures

Method Description V2 / V3
trigger_spectrum_capture(sourceId=-1) Starts a capture. Answer: one spectrum (one per device in V3). sourceId selects the source used by the capture sequence: 0 = none, n = n-th source counted from 1 as in the GUI, -1 = keep the current selection.
trigger_dark_spectrum_capture() Dark capture, stored in SpectroLab. No answer. V2 only
trigger_reference_spectrum_capture(sourceId=-1) Reference capture, stored in SpectroLab. No answer. V2 only
trigger_adc_capture() Reads all ADC channels of the source module (UVx8). Answer: ADC data, read with to_adc_data().

V2 / V3 – In V3, handle dark and reference in the script: capture the dark with the source off, after setting exposure and gain, then subtract it (example 4.2).

3.4 Light sources

set_source(sourceId, isEnabled, pwmValue=-1, dacValue=-1) switches a source or sets its power:

Call Effect
set_source(id, True) / set_source(id, False) Switches the source on / off.
set_source(id, True, pwmValue=p) PWM duty cycle, 0–100 %.
set_source(id, True, dacValue=d) DAC level, 0–100 %.
set_source(id, True, pwmValue=p, dacValue=d) Both.

When pwmValue or dacValue is given, isEnabled is not transmitted: only the power changes. Switch the source explicitly with set_source(id, True) / set_source(id, False).

V2 / V3 – Source index in set_source(): V2 is 1-based, as in the SpectroLab GUI (1 = first source). V3 is 0-based, as in the firmware (0 = first source).

PWM period (V3 ≥ 3.0.9): set_source_pwm_period_us(periodUs, sourceId) sets the period in µs (note the argument order). set_source(..., pwmPeriodUs=p) sends it before the PWM value. The device clamps the value to the source limits (50–1 000 000 µs when the source has no EEPROM).

V2 / V3 – The V2 GUI applies the source’s default PWM period before enabling PWM. In V3, without pwmPeriodUs, the device keeps its current period: pass pwmPeriodUs to reproduce the V2 conditions.

3.5 Several devices (V3)

targetSPNIAddress selects the device:

Constant (SPNIAddress) Value Target
ADDRESS_CURRENT_DEVICE 0x0FF0 Device selected in SpectroLab (default)
ADDRESS_ALL_DEVICES 0x0FFE All connected devices – one answer per device
ADDRESS_BY_DEVICE_OFFSET + i 0x0300 + i Device of index i

Each spectrum carries the address of its device in SPNIAddress. Identify devices by serial number with read_sernum() rather than by index (example 4.2).

3.6 Device values and colorimetry

Method Description V2 / V3
request_read_device_data(keys) Requests one key or a list of keys. Answer: device values. Keys: table below
request_write_device_data(keys, values) Writes one or several uint32 values.
send_pipe_colo_spectrum(name, spectrum, illuminant=D65, degree=DEG_2) Computes colorimetry on a spectrum. Answer: 36 values, read with to_colorimetry_data(). V2 only
Key (SpectrolabDeviceKey) V2 V3
DEV_KEY_BOARD_TEMP Read Read
DEV_KEY_NB_ROWS Write Read, write
DEV_KEY_START_ROW Write Read, write
DEV_KEY_DEV_SN Read (key 4) Read (key 0x0201), or read_sernum()
DEV_KEY_CURR_SRC_NAME Read: selected source name Read: selected source name
Other keys – SPNI model-value keys (communication protocol description, on request)

V3 – Values read from the device come back with type NONE and the SPNI model key. Decode them with to_str(), to_uint() or to_float().

3.7 Received packets

SpectrolabPacket method Returns
opcode() SpectrolabOpcode
to_spectrum() SpectrolabSpectrum: captureType, length_points, X (nm), Y; V3 adds SPNIAddress (source device)
to_adc_data() SpectrolabAdcData: data (list of floats, one per channel)
to_device_data() dict {key: SpectrolabDeviceValue}, with value(), to_uint(), to_float(), to_str()
to_colorimetry_data() SpectrolabColorimetryData: CIE_X, CIE_x, Lab_L, Lab_a, Lab_b, CCT…

4. Examples

4.1 Source in PWM mode

First source at 50 % PWM. Only the source index and the PWM period differ between versions.

# V2: 1-based index, PWM period set by SpectroLab
link.set_source(1, True)
link.set_source(1, True, pwmValue=50)
link.trigger_spectrum_capture()
s = wait_spectrum(link)
link.set_source(1, False)

# V3: 0-based index, explicit PWM period (>= 3.0.9)
link.set_source(0, True)
link.set_source(0, True, pwmValue=50, pwmPeriodUs=1000)
link.trigger_spectrum_capture()
s = wait_spectrum(link)
link.set_source(0, False)

4.2 Several devices with dark subtraction (V3)

# Map serial number -> device address
devices = {}
for i in range(8):
    addr = SPNIAddress.ADDRESS_BY_DEVICE_OFFSET + i
    sn = link.read_sernum(addr)
    if sn:
        devices[sn] = addr

def capture_all():
    link.trigger_spectrum_capture(targetSPNIAddress=SPNIAddress.ADDRESS_ALL_DEVICES)
    spectra = {}
    for _ in devices:
        s = wait_spectrum(link)
        if s:
            spectra[s.SPNIAddress] = s      # address of the source device
    return spectra

for addr in devices.values():               # settings first, then dark
    link.set_expo_ms(100, targetSPNIAddress=addr)
    link.set_gain_mdB(2000, targetSPNIAddress=addr)
    link.set_source(0, False, targetSPNIAddress=addr)
dark = capture_all()

for addr in devices.values():
    link.set_source(0, True, targetSPNIAddress=addr)
    link.set_source(0, True, pwmValue=50, pwmPeriodUs=1000, targetSPNIAddress=addr)
time.sleep(0.5)                             # source stabilisation
light = capture_all()

for sn, addr in devices.items():
    y = np.array(light[addr].Y) - np.array(dark[addr].Y)
    print(sn, y.max())

4.3 Dark, reference and colorimetry (V2)

link.trigger_dark_spectrum_capture()          # stored in SpectroLab, no answer
time.sleep(1)                                 # no completion message: wait
link.trigger_reference_spectrum_capture(1)    # reference with source 1
time.sleep(1)
link.trigger_spectrum_capture(1)
s = wait_spectrum(link)

link.send_pipe_colo_spectrum("sample", s, SpectroloabColorIlluminant.D65)
while True:
    if link.available() > 0:
        pkt = link.read_packet()
        if pkt.opcode() == SpectrolabOpcode.OPCODE_PIPELINE_COLORIMETRY_IN:
            colo = pkt.to_colorimetry_data()
            print(colo.Lab_L, colo.Lab_a, colo.Lab_b, colo.CCT)
            break
    else:
        time.sleep(0.02)

4.4 ADC readout (UVx8 module)

link.trigger_adc_capture()
while True:
    if link.available() > 0:
        pkt = link.read_packet()
        if pkt.opcode() == SpectrolabOpcode.OPCODE_RECEIVED_ALL_CHANNELS_ADC_VALUE:
            print(pkt.to_adc_data().data)    # one value per channel
            break
    else:
        time.sleep(0.02)

5. V2 / V3 summary

Topic V2 (SpectroLab 2.x) V3 (SpectroLab 3.x)
Devices One: the device connected in the GUI Several, addressed individually (3.5)
Source index in set_source() 1-based (GUI numbering) 0-based (firmware numbering)
PWM period Set by SpectroLab set_source_pwm_period_us(), pwmPeriodUs (≥ 3.0.9)
Dark / reference capture Handled by SpectroLab In the script
Colorimetry Yes No
Serial number Key 4 Key 0x0201; read_sernum()
Device value answers Typed (STR, FLOAT…) Type NONE, SPNI model key
read_packet() on empty queue [] None
Helpers – await_packet(), read_sernum()
Spectrum packet 4-byte header 8-byte header, with the source device address

6. Protocol reference

spectroLink.py builds and parses the packets for you. This section is only needed for debugging or for a client written in another language.

6.1 Transport

  • TCP server opened by SpectroLab on 127.0.0.1, port 65357. Several clients can connect.
  • Every packet sent by SpectroLab is broadcast to all clients.
  • No acknowledgement, no CRC, no end key: the receiver resynchronises on the header.
  • Little endian. Floats are IEEE-754 32-bit.

6.2 Frame

Field Bytes Value
Header 6 123, 56, 2, 3, 47, 231 (0x7B 0x38 0x02 0x03 0x2F 0xE7)
Sender address 1 See 6.3. SpectroLink: 7 (ADDRESS_PYLINK)
Number of destinations N 1 SpectroLink: 1
Destination addresses N SpectroLink commands: 1 (ADDRESS_ALL_DEVICES)
Opcode 1 See 6.4
Payload length L 2 uint16
Payload L See 6.4 – 65 535 bytes maximum

6.3 Addresses

Value Intercom address Value Intercom address
0 ADDRESS_ALL 4 ADDRESS_LIGHTING
1 ADDRESS_ALL_DEVICES 5 ADDRESS_EEPROM
2 ADDRESS_SPECTRUM 6 ADDRESS_TIMELAPSE
3 ADDRESS_BEER_LAMBERT 7 ADDRESS_PYLINK

In V3, every command payload starts with the 4-byte target device address (uint32, see 3.5).

V2 / V3 – V2 payloads have no address prefix. A V2 client on V3, or the reverse, shifts every field.

6.4 Opcodes

Direction: → script to SpectroLab, ← SpectroLab to script. Payloads are given after the V3 address prefix.

Code Name (OPCODE_…) Dir. Payload V2 V3
1 SET_EXPO_MS → float exposure (ms) Yes Yes
2 SET_GAIN_MDB → float gain (mdB) Yes Yes
3 TRIGGER_SPECTRUM_CAPTURE → Source (1): V2 optional uint8, V3 uint32 Yes Yes
5 TRIGGER_ALL_CHANNELS_ADC_CAPTURE → – Yes Yes
6 / 7 SET_SOURCE_ON / OFF → uint8 source Yes Yes
8 SET_SOURCE_PWM → uint8 source, uint8 PWM % Yes Yes
15 SET_SOURCE_DAC → uint8 source, uint8 DAC % Yes Yes
16 SET_SOURCE_PWM_AND_DAC → uint8 source, uint8 PWM %, uint8 DAC % Yes Yes
23 SET_SOURCE_PWM_PERIOD → uint8 source, uint32 period (µs) No ≥ 3.0.9
13 REQUEST_READ_DEVICE_VALUE → uint8 n, n × uint32 key Yes Yes
14 REQUEST_WRITE_DEVICE_VALUE → uint8 n, n × uint32 key, n × uint32 value Yes Yes
19 TRIGGER_DARK_SPECTRUM_CAPTURE → – Yes No
20 TRIGGER_REF_SPECTRUM_CAPTURE → optional uint8 source (1) Yes No
22 PIPELINE_COLORIMETRY_OUT → See 6.7 Yes No
9 RECEIVED_SPECTRUM ← See 6.5 Yes Yes
11 RECEIVED_ALL_CHANNELS_ADC_VALUE ← uint8 n, uint8 size (4), n × float Yes Yes
12 RECEIVED_DEVICE_VALUE ← See 6.6 Yes Yes
21 PIPELINE_COLORIMETRY_IN ← uint16 name length, name, uint16 n (36), n × float Yes No
17 / 18 PIPELINE_MATH_SPEC_IN / OUT ↔ Internal math pipeline (internallink.py) Int. Int.

(1) 0 = none, n = n-th source (1-based); absent (V2) or 0xFFFFFFFF (V3) = keep the current selection. Codes 4 and 10 are reserved. “No”: the packet is ignored.

6.5 Spectrum payload

Field Bytes Description
Source device address (V3 only) 4 Address of the device that produced the spectrum
Capture type 1 2 = wavelength (nm)
Number of points N 2 uint16
Bytes per point 1 4 (float32)
X 4·N Wavelengths (nm)
Y 4·N Intensities

Only standard captures are returned. Dark and reference captures (V2) update SpectroLab’s internal dark and reference and produce no packet.

6.6 Device value payload

Field Bytes Description
n 1 Number of values
Keys 4·n uint32 keys
Types n 0 NONE, 1 UINT8, 2 UINT16, 3 UINT32, 4 FLOAT, 5 STR
Lengths 4·n uint32, length in bytes of STR / NONE values
Values Σ 1, 2 or 4 bytes for numeric types, “length” bytes for STR / NONE

Key values (SpectrolabDeviceKey): BOARD_TEMP 1, NB_ROWS 2, START_ROW 3, DEV_SN 4 (V2) / 513 (V3), CURR_SRC_NAME 5.

6.7 Colorimetry request (V2)

Field Bytes Description
Name length 2 uint16
Name var. UTF-8, returned with the result
Illuminant 1 A 0, B 1, C 2, D50 3, D65 4, E 5
Observer 1 0 = 2°
Capture type 1 2 (nm)
N, bytes per point 2 + 1 uint16, 4
X, Y 4·N each float32

The answer (PIPELINE_COLORIMETRY_IN) holds 36 floats, in this order: 15 calculation coefficients (xr … vppr), CIE XYZ, CIE xy, L*a*b*, RGB, LCHab, L*u*v*, LCHuv, CCT.


Need help with an integration, or the full device-key description for V3? Contact our R&D team. SpectroLab and SpectroLink are free, with unlimited licences: download SpectroLab.