Optical Filter Characterization

The IndiGo UV/Vis or the IndiGo NIR, combined with the GoLight, make a powerful, simple and effective tool for characterizing optical filters in the visible and near infrared.

They determine the passbands, the optical density (OD) and the cut-off slope(s). In a few seconds, you can identify the type of filter (band-pass, long-pass, edge, dichroic…).

It is a reliable, accurate and affordable way to qualify a laboratory filter, for example one that was mixed up after not being returned to its original packaging!

The GoLight provides a continuous white light spectrum from 380 nm to 950 nm. An optical filter adapter is placed on the GoLight. The adapter is designed for round filters of 0.5, 1 or 2 inches; other sizes can be made to order. Contact us.

The principle is simple: in SpectroLab, take the white light of the GoLight as a reference, place your filter in the filter adapter and capture the spectrum. You instantly get the transmission (or absorption) spectrum of your optical filter.

Figure 1 below shows characterization tests of Kodak color filters with this setup: the measured absorption spectra closely match the theoretical spectra.

Kodak optical filter characterization with the IndiGo portable visible spectrometer

Kodak optical filter characterization with the IndiGo portable visible spectrometer